Product Services

Description

Measurement Principle of the Hydrogen Probe /Hydrogen determination probe
Precise control of trace inert gases is employed; the gas is introduced via a submerged hydrogen‑sensing probe’s delivery tube into a ceramic chamber containing a molten steel sample, allowing equilibrium to be established between the gas and the steel. A vacuum pump then draws the equilibrated gas into the sensing device.
The hydrogen‑determination probe provides an interface for inert gas and for hydrogen in the molten steel once equilibrium has been reached. A small tube connects the probe to the hydrogen analyzer, drawing gas into the system’s main line to deliver an adequate flow to the probe. Hydrogen is transferred very rapidly through the probe’s interface. The balance gas is drawn by a vacuum pump into the instrument’s sensing unit and then vented to the atmosphere, forming an open loop. The sensing unit measures thermal conductivity to determine the partial pressure of hydrogen, from which the hydrogen concentration is calculated using an instrument‑specific formula.


Hydrogen Analyzer Technical Specifications /Technical parameter of the hydrogen determination probe
1. Measurement range: 0.3–9.9 ppm
2. Repeatability: ±5%
3. Measurement cycle: 0.4 to 2 minutes
4. Hardware:
Displays 3½ digits, 1-inch-high LED graphic waveform display (when connected to a computer), and waveform printing capability.
Communication: RS232 (within 15 m) or RS422 (within 1 km)
Computer: Windows 98 or later (for data storage)
5. Source of Supply:
Nitrogen: 99.995% purity, 40 mL, 12 PSIG; Compressed air: 80–200 PSIG.
Power supply: 10 VA, 100–240 VAC, 50–60 Hz
6. Mode:
Measurement mode; Adjustment mode; Calibration mode
7. Chronological order: Normal mode: 30–60 seconds
8. Measurement Status Display
Yellow Light: System purging and automatic zeroing
Green light: Probe immersed in molten steel for purging and measurement.
Red light/Warning sound: Test completed


Product Specifications and Packaging /Spec of the product and Packing

Product model Paper tube specifications Carton dimensions (L×W×H) mm Number of interior fittings (pieces) Total Weight (KG)
H900 Φ60×1055 185×125×1010 6 4.5

Keywords:

Online Consultation